System S380 Dual-Channel Laboratory Photometer
Manual | Specifications
The System S380 has all of the System S370s convenient features: easy programming, an IEEE-488 interface for host computer control, and storage capacity for a large number of sensor head calibrations.
However, an additional measurement channel is provided with the S380. So it is simple to measure two light sources simultaneously, or perform real-time ratio measurements.
The second channel is as easy to program as the first, since it is prompted and controlled similarly. And the prompt menus and measurement data are displayed on a large, 32-character LCD screen.
Performance Specifications
|
Linear Mode |
Log Mode |
|
Accuracy
|
|
|
| 103 to 107 range |
+1 count
|
+1 count
|
| 108 to 109 range |
+1 count
|
+1 count
|
|
Dynamic Range
|
10 decades
|
9 decades
|
|
Calibration
|
May be calibrated
in multiple units
|
|
|
Temperature Drift
|
|
103 to 107 range
|
±0.03% /Cº
|
N/A
|
|
108 to 109 range
|
±0.06% /Cº
|
N/A
|
dBm, percent of reading,
referred to input
|
N/A
|
None
|
General Specifications
|
Display
|
Liquid crystal, with two lines of 16 characters each
|
|
Display Range
|
0000 to 9999
|
|
Range Control
|
Automatic or manual
|
|
Output
|
|
Digital
|
IEEE-488 (drivers available for Lab Windows)
|
|
Analog
|
0-2.5 VDC
|
|
Output Connectors
|
|
Digital
|
IEEE-488
|
|
Analog
|
BNC
|
|
Bandwidth
|
9.0 Hz
|
|
Temperature Range
|
|
Operating
|
0º to +45ºC
|
|
Storage
|
-20º to +60ºC
|
|
Weight
|
4.0 pounds (1.8 kg)
|
|
Accessories Supplied
|
Power cord, instruction/maintenance manual
|
|